Authors
Suetov A.A.1, Boiko E.V.1, 2, 3, Doktorova Т.А.1, Izmaylov A.S.1
1 St. Petersburg Branch S. Fyodorov Eye Microsurgery Federal State Institution, St. Petersburg
2 Mechnikov North-West State Medical University, St. Petersburg
3 St. Petersburg State University, St. Petersburg
Abstract
Background. Modern laser systems utilize patterned delivery technology to speed up the treatment process. However, the pattern-based application of burns does not take into account potential variations in the intensity of burns formed at individual points due to differing treatment conditions (such as differences in fundus pigmentation, optic media transparency, and changes in focus distances). The extent to which such variations occur in patterned burns is not yet fully understood.
Aim. To investigate the homogeneity of burn intensity on the fundus during pattern laser coagulation performed under conventional settings.
Materials and methods. Fundus images of 4 eyes were analyzed using ImageJ software following pattern laser photocoagulation performed through a pan-fundus lens, with burns applied in 90 patterns consisting of 7 points located in three zones: Zone 1, corresponding to a 450-field centered on the fovea, Zone 2, corresponding to a field ranging from 450 to 900, and Zone 3, covering areas of the fundus more peripheral than Zone 2. The following parameters were used for the laser treatment: a wavelength of 532 nm, a retinal spot diameter of 400 μm, pulse durations of 20 and 30 ms, and a target intensity of degree 2 according to F.A. L’Esperance. The intensity and coefficient of variation of the burns in the patterns across different zones were analyzed.
Results. The burn intensity in Zone 1 was significantly higher for 30-ms pulses than for 20-ms pulses (1.13±0.03 and 1.09±0.04, p < 0.001). In zone 2, the burn intensity for pulse durations of 20 ms and 30 ms was 1.17±0.04 and 1.18±0.08 (p = 0.08), and in Zone 3, 1.18±0.04 and 1.19±0.08, respectively (p = 0.25). Burn intensity differed significantly across zones both when using 20-ms pulses (F = 143.82, p < 0.001) and when using 30-ms pulses (F = 20.92, p < 0.001). The coefficient of variation for burn intensity in the patterns was 1.93±0.78% for 20-ms pulse durations and 2.27±1.17% for 30-ms pulse durations (p = 0.127), with a significant difference in this indicator observed in Zone 1 (1.42±0.42% and 2.09±0.88%, p = 0.013).
Conclusion. The use of fixed treatment parameters in pattern laser coagulation may result in burns of varying intensity, both within separate patterns and in areas of the fundus located at different distances from the posterior pole. A more pronounced difference in burn intensity is observed when using 30-ms pulses compared to 20-ms pulses.
Keywords: retina, laser, laser photocoagulation, pattern, burn intensity.
References
1. Grzybowski A, Luttrull JK, Kozak I, editors. Retina Lasers in Ophthalmology. Cham: Springer International Publishing; 2023. doi: 10.1007/978-3-031-25779-7.
2. Boiko EV. Lazery v oftal’mohirurgii: teoreticheskie i prakticheskie osnovy. SPb: VMedA im. S.M. Kirova; 2003. (In Russ.)
3. Izmaylov AS, Balashevich LI. Panretinal photocoagulation in diabetic retinopathy: indications for additional treatment. Ophthalmol Reports. 2009; 2: 44–51. (In Russ.)
4. Chauhan DS, Downie JA, Eckstein M, Aylward GW. Failure of prophylactic retinopexy in fellow eyes without a posterior vitreous detachment. Arch Ophthalmol. 2006; 124: 968–71. doi: 10.1001/archopht.124.7.968.
5. Muqit MMK, Gray JCB, Marcellino GR, et al. In vivo laser-tissue interactions and healing responses from 20- vs 100-millisecond pulse pascal photocoagulation burns. Arch Ophthalmol. 2010; 128: 448–55. doi: 10.1001/archophthalmol.2010.36.
6. Muqit MMK, Denniss J, Nourrit V, et al. Spatial and spectral imaging of retinal laser photocoagulation burns. Investig Ophthalmol Vis Sci. 2011; 52: 994–1002. doi: 10.1167/iovs.10-6309.
7. L´Esperance FA. Ophthalmic Lasers. Photocoagulation, Photoradiation and Surgery. St. Louis: Mosby; 1989.
8. Lanzetta P, Dorin G, Pirracchio A, Bandello F. Theoretical bases of non-ophthalmoscopically visible endpoint photocoagulation. Semin Ophthalmol. 2001; 16: 8–11. doi: 10.1076/soph.16.1.8.4216.