DOI: 10.25881/20728255_2026_21_4_S1_57

Authors

Makov I.N.1, Kornilovskiy I.M.2

1 Astana Vision, Almaty, Republic of Kazakhstan

2 Pirogov National Medical and Surgical Center, Moscow

Abstract

Aim. To assess whether separate modification of the front and back optical-zone asphericity can control optical-zone edge thickness, induced spherical aberration, and peripheral tear clearance.

Materials and methods. Using the example of a lens with a refractive power of −8.0 D, a comparison was made between a spherical minus lens, a lens with an aspheric front surface, and a biaspheric lens. The front surface was modified to reduce thickness at the optical-zone edge by 5-40 μm. Induced spherical aberration was calculated for a 6-mm pupil. The back surface was then modified until complete compensation, a 50-μm reduction in peripheral clearance, or loss of the thickness advantage. For thinning values of 10, 20, 30, and 40 μm, a pupil-matched zonal back asphere was also calculated for pupil diameters of 3-6 mm and blended to the original profile over a 0.25-mm transition.

Results. The thickness of the spherical lens at the optical-zone edge was 376.6 μm. When the back asphere extended over the full 8-mm optical zone, complete compensation made the lens thicker than the spherical design, and no more than 27.7% of the aberration could be compensated while retaining a thickness advantage. Pupil-matched zonal compensation for 3-5-mm pupils retained the prescribed thinning throughout the 10-40-μm range. For a 6-mm pupil, the full 10-μm thinning was retained; 16.0 and 2.4 μm remained from the prescribed 20- and 30-μm thinning, respectively, whereas a 40-μm target increased the maximum thickness to 9.6 μm above that of the spherical design.

Conclusions. Separate modification of the front and back surface asphericity can be used in the analytical design of a scleral lens optical zone. In the investigated minus-powered model, a full-diameter back asphere eliminated the edge-thickness advantage when spherical aberration was fully compensated. Limiting the back asphere to the pupil and adding a short transition reduced its effect on maximum thickness and peripheral clearance. The prescribed thinning of up to 40 μm was retained for 3-5-mm pupils; for a 6-mm pupil, the advantage was retained only with a small thinning target.

Keywords: scleral contact lenses, analytical design, optical zone asphericity, spherical aberration, tear clearance, lens thickness.

References

1. Sabesan R, Johns L, Tomashevskaya O, et al. Wavefront-guided scleral lens prosthetic device for keratoconus. Optom Vis Sci. 2013; 90(4): 314-323. doi: 10.1097/OPX.0b013e318288d19c.

2. Badrinarayanan A, Balakrishnan AC, Dutta R, et al. Impact of scleral lens front surface eccentricity on visual acuity, contrast sensitivity, and higher-order aberrations in eyes with keratoconus. Eye Contact Lens. 2023; 49(9): 374-378. doi: 10.1097/ICL.0000000000001007.

3. Kumar M, Shetty R, Dutta D, et al. Effects of a semi-scleral contact lens on refraction and higher order aberrations. Cont Lens Anterior Eye. 2019; 42(6): 670-674. doi: 10.1016/j.clae.2019.06.002.

4. Domínguez-Vicent A, Esteve-Taboada JJ, Recchioni A, Brautaset R. Power profiles and in vitro optical quality of scleral contact lenses: effect of the aperture and power. Eye Contact Lens. 2018; 44(3): 149-158. doi: 10.1097/ICL.0000000000000345.

5. Lindskoog Pettersson A, Mårtensson L, Salkic J, et al. Spherical aberration in relation to visual performance in contact lens wear. Cont Lens Anterior Eye. 2011; 34(1): 12-16. doi: 10.1016/j.clae.2010.08.008.

6. Bakalova NA. “OKVision® SMARTFIT™” Customizable Mini-Scleral Lenses Fitting Guide. The EYE GLAZ. 2021; 23(1): 41-51. (In Russ). doi: 10.33791/2222-4408-2021-1-41-51.

7. Patent RUS №2858071/11.03.2026. Makov IN. Zhestkaya kontaktnaya linza s vnutrennei poverkhnost’yu, sostoyashchei iz zon, obrazovannykh vrashcheniem konicheskikh sechenii. https://patents.google.com/patent/RU2858071C1/ru. (In Russ.)

For citation

Makov I.N., Kornilovskiy I.M. Analytical design of a biaspheric optical zone in a scleral contact lens. Bulletin of Pirogov National Medical & Surgical Center. 2026;21(4,supplement):57-62. (In Russ.) https://doi.org/10.25881/20728255_2026_21_4_S1_57